User Manual

Operation Cond 1970i
46
ba75692e01 01/2007
4.7 Configuration
You can adapt the measuring instrument to your individual require-
ments. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Note
You can leave the configuration menu at any time with <M> . The pa-
rameters that have already been changed are stored.
Baud rate
Baud rate 1200, 2400, 4800, 9600
Calibration interval (Int 3) 1 ... 180 ... 999 d
AutoRange ARng On or off
Reference temperature 25 °C (TREF25)
20 °C (TREF20)
Date/time Any
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
4 Select the required baud rate with <> <>.
5 Confirm with <RUN/ENTER>. Int 3 appears on the display.
REL1
Store