Display LoBat
Instrument does not
react to keystroke
Error message OFL
OFLOFL
OFL
Error message E3
E3E3
E3
Display to
toto
to
Cause Remedy
Battery almost empty Charge the battery (see
section 3.2)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <RCL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause Remedy
The measured value lies outside
the measuring range
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and
replace it if necessary
Unsuitable calibration
solution
Check calibration solutions
Cause Remedy
Time-out of the interface Check the instrument
connected