User Manual

Cond 1970i Technical data
59
ba75692e01 01/2007
Accuracy
(± 1 digit)
Variable Accuracy at test sample
temperature
, temperature compensation (TC)
none ± 0.5 %
Nonlinear ± 0.5 % 0 °C ... + 35 °C
according to EN 27 888
± 0.5 % + 35 °C ... + 50 °C
Extended nLF function
according to WTW mea-
surements
Linear ± 0.5 % + 10 °C ... + 75 °C
SAL / Range
0.0 ... 42.0 ± 0.1 + 5 °C ... + 25 °C
± 0.2 + 25 °C ... + 30 °C
TDS [mg/l]
± 1
T [°C] / Temperature sensor
NTC 30 ± 0.1 0 °C ... + 55 °C
PT 1000 ± 0.5
± 0.1
± 0.5
0 °C ... + 15 °C
+ 15 °C ... + 35 °C
+ 35 °C ... + 55 °C