User guide
Cond 3210 Technical data
49
ba75794e01 01/2009
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the measuring cell has also to be taken into account.
Accuracy (± 1 digit) Parameter Accuracy Temperature of the test
sample
ӕ and
ρ
/ temperature compensation
None (Off) ± 0.5 %
Nonlinear (nLF) ± 0.5 % 0 °C ... + 35 °C
according to EN 27 888
± 0.5 % + 35 °C ... + 50 °C
enhanced nLF function
Linear (lin) ± 0.5 % + 10 °C ... + 75 °C
SAL / range
0.0 ... 42.0 ± 0.1 + 5 °C ... + 25 °C
± 0.2 + 25 °C ... + 30 °C
TDS [mg/l]
± 0.5 %
T [°C] / temperature sensor
NTC 30 ± 0.2
PT 1000 ± 0.3