Instruction Manual

Maintenance, cleaning, replacement ConOx
34
4.3 Storage
Always store the sensor with the DO module in the calibration vessel at a
temperature of 0 to +50 °C. Make sure that the sponge in the calibration ves-
sel is always moist.
Storage in the
calibration vessel,
OxiCal
®
-CX
Moisten the sponge:
l Remove the cap (1).
l Take out the sponge (2), wet it, then slightly squeeze it out.
l Reinsert the sponge and close the calibration vessel with the cap.
5 Maintenance, cleaning, replacement
5.1 General maintenance instructions
For your safety Note the following safety instructions when handling electrolyte and cleaning
solutions:
Caution
The ELY/G electrolyte solution irritates eyes and skin. When dealing
with the ELY/G electrolyte solution, observe the following points:
l During working activities, always wear suitable protective gloves
and protective goggles/face shield.
l If it comes into contact with the skin, rinse thoroughly with water and
immediately change contaminated clothing.
l If it comes into contact with the eyes, rinse thoroughly with water
and consult a doctor.
l Follow the safety datasheet.
12