Instruction Manual

What to do if... ConOx
42
6 What to do if...
6.1 Error symptoms of DO measurement
6.2 Error symptoms of conductivity measurement
Error symptom Cause Remedy
The sensor is in the air
and the display shows
0.0 mg/l or 0 % O
2
No connection between
measuring instrument and sensor
Cable defective
No electrolyte in the membrane
cap
Check connection between
measuring instrument and sensor
Replace and refill the membrane
cap (see section 5.3)
The sensor cannot be
calibrated
Contaminated membrane cap
Electrolyte depleted
Replace and refill the membrane
cap (see section 5.3)
Subsequently, wait for 30 to
50 min and recalibrate.
The sensor still cannot
be calibrated after
changing the electrolyte
and membrane cap
Contaminated electrodes or
sensor toxification
membrane cap not screwed on
tight enough
Clean the electrodes (see
section 5.4)
Screw membrane cap tighter
Incorrect temperature
display
The temperature sensor of the
conductivity basic module was not
immersed deep enough in the
measuring solution
Temperature sensor defective
Observe the minimum immersion
depth
Return the sensor
Mechanical damage to
the sensor
Return the sensor
Error symptom Cause Remedy
No temperature or con-
ductivity display
No connection between
measuring instrument and sensor
Cable defective
Check connection between
measuring instrument and sensor
Measurement delivers
implausible conductivity
values
Incorrect cell constant adjusted at
the measuring instrument
Measuring range exceeded
Contamination in the area of the
electrodes
Electrodes damaged
Check / correct the cell constant
Observe the application range
Clean the sensor (see
section 5.2).
Return the sensor
Incorrect temperature
display
The temperature sensor was not
immersed deep enough in the
measuring solution
Temperature sensor defective
Observe the minimum immersion
depth
Return the sensor