User Manual

ba80000def01_EBI16 13
English
Français
EBI16 checking the B&D test
Starting conditions
- EBI16 must be started at room
temperature (+15°C +35°C).
Place in the chamber
- If necessary, preheat the sterilizer
(sterilizer manual)
- Place the EBI16 in the center of the
chamber.
- Place the EBI16 vertical in the
chamber.
- Make sure that the EBI16 is not in the
area of the steam entry.
- Start the sterilizer immediately after
placement of the EBI16
- During the B&D test, no other load is
allowed in the chamber.
Evaluation
- Interface and software are required
for the operation of the EBI16
- Automatic evaluation of the B&D test
can only be carried out with
Winlog.med or Winlog.validation.
Archiving of the results
- The archiving of the results is
ensured permanently by Winlog.med
or Winlog.validation.
Effectuer le test de B&D avec
l’EBI 16
Conditions de depart
- L’EBI 16 doit être démarré à
température ambiante
(+15°C…+35°C).
Positionnement dans la chambre
- Si nécessaire, préchauffer le
stérilisateur
- Positionner l’EBI 16 au centre de la
chambre
- Positionner l’EBI 16 verticalement
dans la chambre
- Veuillez vous assurer que l’EBI 16 ne
se trouve pas à l’endroit où pénètre la
vapeur.
- Démarrer le stérilisateur
immédiatement après avoir
positionné l’EBI 16
- Pendant le test de B&D, aucune autre
charge n’est autoriseé dans la
chambre
Evaluation
- L’interface et le logiciel sont
nécessaires pour utiliser l’EBI 16
- L’évaluation automatique du test de
B&D peut être effectuée seulement
avec les logiciels Winlog.med ou
Winlog.validation.
Archivage des résultats
- L’archivage des résultats est assuré
en permanence par le logiciel
Winlog.med ou Winlog.validation