4 What to do if...
Error symptom Cause Remedy
No temperature value or
D.O. value
No connection between meter
and D.O. sensor
Establish connection between
meter and D.O. sensor
Cable defective Return the D.O. sensor
Measured value too
high or too low
or
error message,
Error
Coating on sensor cap Clean the outside of the sensor
(see section 3.3.1)
Membrane damaged Exchange the sensor cap
Service life of the sensor cap
over
Check the sensor
Replace the sensor cap as
necessary (see section 3.2)
Dirt inside the sensor cap and
sensor head
Fixing ring not properly tightened
Sensor cap untight or defective
Dismantle the sensor cap
Clean the insides of the sensor
cap and sensor head (see
section 3.3.2)
Mount the sensor cap correctly
and tighten the fixing ring as far
as it will go (see section 3.2)
Replace a defective sensor cap
as necessary
Incorrect temperature
display
The temperature sensor is not
immersed deep enough in the
measuring solution
Observe the minimum immersion
depth
Temperature sensor defective Return the D.O. sensor
Error message,
no cap
No sensor cap put on Clean the sensor head and
sensor cap
Sensor cap not recognized Return the D.O. sensor
Sensor cap defective Replace the sensor cap