Owner manual

Global Water
800-876-1172 globalw.com
- 5 -
Level Sensor:
Sensor Element: Silicone Diaphragm, Wet/Wet Transducer
Range: Level: 0-3 feet
Temperature: 32°-122°F (0°-50°C)
Linearity and Hysteresis: ±0.1% FS
Accuracy: Level: ±0.1% FS at constant temperature
±0.2% over 32° to 70°F range
Temperature: Smaller of 0.5°F or ±1% of reading
Overpressure: 2 x full scale
Burst Pressure: 10 x full scale
Resolution: Infinite (Analog)
Outputs: 4-20mA ±1mA at full scale
Optional Temperature: 0-10mA ±1mA FS
Supply Voltage: 10-36VDC
Current Draw: Sum of sensor outputs.
Warm Up Time: 10mS Min, 3 sec. recommended
Operating Temperature: (Not Frozen) to +185°F
Compensated Range: 32° to 70°F submerged, automatic barometric
compensation
Sensor Housing:
Material: 316 SS
Size: 9” long x 1.0” diameter
Weight: 250g (9oz)
Cable:
Conductors: 4 each 22 AWG
Jacket Material: 87A shore hardness Polyurethane
Optional jacket: Fluorinated Ethylene Propylene (FEP) Teflon
Cable O.D.: 7.8mm (0.307”)
Vent tube: HD Polyethylene
Shield: Aluminum Mylar
Temperature range: -30 to 85°C (-22 to 185°F)
Weight: ~65g/m (0.7 oz/ft)