Specifications
Range 0.10 to 9.99ppm (mg/l)
Accuracy ± 3% of reading or ± 0.1ppm (whichever is greater)
Resolution 0.1ppm
Display 2000 count, Dual function 3 ½ digit LCD with Bargraph,
Display size: 24 mm x 20 mm
Electrode Europium doped lanthanum fluoride single crystal
Electrode life 6 months minimum
Response Time 90% of change in less than 30 seconds (typical)
Operating Temp. Range 32 to 140°F (0 to 60°C)
ATC Range 32 to 140°F (0 to 60°C)
Measurement Storage 25 tagged (numbered) data sets with recall
Battery Power Four (4) CR2032 button batteries
Low Battery Indication ‘BAT’ appears on the LCD
Auto Power Off After 12 minutes of inactivity
Dimensions/Weight 1.4 x 6.8 x 1.6” (36 x 173 x 41mm); 7.4 oz (210g)