What to do if... HandyLab 100
56
ba77062d01 03/2014
No stable measured
value
Sensor symbol flashes
Display
Cause Remedy
pH electrode:
Junction contaminated Clean the junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded
if necessary
Temperature not stable Adjust temperature if nec-
essary
Electrode + test sample:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Batteries almost empty Replace the batteries (see
section 5.1 M
AINTENANCE)