Owner's manual
Operation HandyLab 200
28
ba77063e01 03/2014
Nonlinear temperature compensation (nLF) according to EN 27
888
Linear temperature compensation (lin) with adjustable coeffi-
cients of 0.000 ... 3.000 %
/K
No temperature compensation (off)
Note
The reference temperature and temperature compensation are set in
the Measurement menu for the parameter, conductivity (see section
4.4.4).
Application tips Select the following temperature compensations given in the table
according to the respective test sample:
Test sample Temperature compensation Display
indica-
tor
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water nLF
according to EN 27 888
nLF
Other aqueous solu-
tions
lin Set linear
temperature coefficient
0.001 ... 3.000 %/K
lin
Salinity (seawater) Automatic nLF according to
IOT (International Oceano-
graphic Tables)
Sal, nLF