Owner's manual
HandyLab 200 Operation
29
ba77063e01 03/2014
4.4.4 Settings for conductivity measuring cells
Overview The following settings are possible for conductivity measuring cells:
Calibration record (display)
Calibration interval
Measuring cell/cell constant
Reference temperature
Temperature compensation
TDS factor
Unit of the temperature
Automatic Stability control
Settings The settings are made in the Measurement menu for the measured
parameter, conductivity. To open the settings, display the required
parameter in the measured value display and press the <F1>/[menu] or
<ENTER> key. After completing the settings, switch to the measured
value display with <M>.
Menu item Possible set-
ting
Explanation
Calibration /
Calibration record
- Displays the calibration
record of the last calibra-
tion.
Calibration /
Calibration interval
1 ... 999 d Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.