Owner's manual

Operation HandyLab 200
38
ba77063e01 03/2014
4.6 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the
default settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, acti-
vate the relevant measuring window in the measured value display and
press the <F1>[Menu] key shortly
.
Setting Default settings
Cal. interval 150 d
Measured parameter
ϰ
Cell constant (c) calibrated: 0.475 cm
-1
adjusted: 0.475 cm
-1
calibrated: cm
-1
adjusted:
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature compen-
sation
2.000 %/K
TDS factor 1.00
Stability control On
Temperature unit °C