Calibration record
(USB output)
-
-30 ... <-25
or
->+25 ... +30
-62.0 ... <-61.0
or
>-56.0 ... -50.0
Clean the IDS sensor according to the sensor operating manual
Error Error
<-30
or
>+30
<-62.0
or
> -50,0
Error elimination (see section 14 W
HAT TO DO IF..., page 88)
For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.8 QSC
FUNCTION
(
SENSOR QUALITY CONTROL), page 34).
Display Calibration
record
Zero point
[mV]
Slope [mV/pH]
CALIBRATIONpH
01.02.2014 07:43:33
BlueLine 24 pH IDS
Ser. no. B092500013
Buffer 1
Buffer 2 7.00
Buffer 3
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope -60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++