User guide

What to do if... HandyLab 600
60 ba77064e01 03/2014
No stable measured
value
Obviously incorrect
measured values
11.2 General information
Sensor symbol
flashes
Display
Cause Remedy
IDS pH sensor:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if neces-
sary
Temperature not stable Adjust temperature if necessary
IDS pH sensor + test sample:
Conductivity too low Use suitable IDS pH sensor
Temperature too high Use suitable IDS pH sensor
Organic liquids Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor:
Sensor unsuitable Use suitable IDS sensor
Temperature difference
between buffer and test sample
too great
Adjust temperature of buffer or sam-
ple solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Calibration interval expired Recalibrate the measuring system
Cause Remedy
Batteries almost empty Replace the batteries (see section
10.1 M
AINTENANCE, page 57)