Instruction Manual

37
37
Safety guidelines (HandyLab 750 EX)
Warning!
Only open the battery compartment of the HandyLab 750, 750
EX outside the hazardous location.
Never try to open the device. If a repair should be required,
return the device to our factory.
Never use the USB port within the hazardous location.
Warning!
When using the HandyLab 750 EX (device variant for applica-
tions in hazardous locations) in a hazardous location, only the
battery types listed below may be used. The batteries must be
from the same manufacturer and of identical type and capaci-
ty. Never use new and used batteries together.
Batteries for Application in Hazardous Locations
Batteries (4x each) Temp. class Ambient temperature range
Duracell MN1500 T4 -10 °C ≤ Ta ≤ +40 °C
Energizer E91 T3 -10 °C ≤ Ta ≤ +50 °C
Power One 4106 T3 -10 °C ≤ Ta ≤ +50 °C
Panasonic Pro Power LR6 T3 -10 °C ≤ Ta ≤ +50 °C
Precautions for application in hazardous locations
In hazardous areas, only the HandyLab 750 EX may be used. The following
safety guidelines must be observed: