Data Logger
The Data Logger
The meter provides a data logger. Prior to use, it must be confi gured and
then activated. You can choose from the following logger types:
DIFF (signal-controlled logging of measured variable and temperature)
INT (time-controlled logging at a fi xed interval)
DIFF+INT (combined time- and signal-controlled logging)
SHOT (manual logging by pressing the STO key)
The data logger records up to 5000 entries and saves them in a circular buf-
fer.
Already existing entries will be overwritten.
The following data are recorded: primary value, temperature, time stamp and
device status.
The HandyLab Pilotsoftware allows convenient management of the data
logger. It is always the currently selected process variable which is recorded.
The “STO” icon and the memory address is displayed briefl y to indicate that
an entry is being saved.