Manual
54
pH
Confi guration
pH Confi guration
1) In measuring mode, press the Menu softkey.
2) Select “Confi guration” and confi rm by pressing enter.
3) Make the desired adjustments.
The following table gives you an overview.
Factory settings are shown in bold print.
*) “+” indicates that submenus can be opened by pressing enter.
**) Parameter can be confi gured using HandyLab Pilot software.
“Confi guration” menu selection – part 1
Language Deutsch | English | Español | Italiano | Français | Portu-
guês
Auto-off Off | 5 min | 10 min | 30 min | 60 min
Temperature °C | °F
+ pH sensor
*)
Wear On | Off
+ Calibration
*)
Calibration timer Off | On
Interval On: 00 … 99 days
Calibration mode AutoC
al | Manual | Data entry
Calibration points Auto | 1-point | 2-point | 3-point
Buffer set
Mettler-Toledo
Knick CaliMat
Ciba
NIST technical
NIST standard
Hach
WTW
Hamilton
Reagecon
DIN 19267
User buffer 1
**)
.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
MemoLog Off | On
TAG Off | On
-01-
(De-
fault)
NIST standard
-02- NIST technical
-03-
SI Analytics
technical
-04- Knick Calimat
-05- Mettler Toledo
-06- HACH
-07- Ciba (94)
-08- Hamilton
-09- Reagecon
-10- DIN 19267
1,679/4,006/6,865/9,180
1,68/4,00/7,00/10,01/12,46
2,00/4,00/7,00/10,00
2,00/4,00/7,00/9,00/12,00
2,00/4,01/7,00/9,21
4,01/7,00/10,01/12,00
2,06/4,00/7,00/10,00
2,00/4,01/7,00/10,01/12,00
2,00/4,00/7,00/9,00/12,00
1,09/4,65/6,79/9,23/12,75
User-Puffer 1
**)