4.2.3 Salinity
You can measure the salinity as follows:
4.2.4 TDS (Total dissolved solids)
You can measure the total dissolved solids as follows:
1 Perform the preparatory activities according to section 4.2.1.
2 Immerse the conductivity measuring cell in the test sample.
3 Press the <M> key until Sal appears on the status display. The sa-
linity value appears on the display.
4 Wait for a stable measured value.
Tref25
/
K
cm
1/
°
F
%
8
mg/l
1
RCL
ARng
AR
LoBat
Cal
Tref20
Lin
m
S/
cm
M
TDS
S
Auto
Store
Sal
mbar
%
OUpH
mV/pH
cm
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Ident
Year
2
1
2
2
3
3
°
TP
Sal
C
nLF
1 Perform the preparatory activities according to section 4.2.1.
2 Immerse the conductivity measuring cell in the test sample.
3 Press the <M> key until TDS appears on the status display.
The value of the total dissolved solids appears in the upper display
line. The TDS factor appears in the lower display line.
4 Using <V> <W>, set the TDS factor (0.40 ... 1.00).
(The TDS factor has to be determined by a comparison measure-
ment before.)