4.3.2 Setting the cell constant manually
Note
The cell constant to be set must either be taken from the operating manual of
the measuring cell or is printed on the measuring cell.
You can set the cell constant manually as follows:
Range
0.250 ... 2.500 cm
-1
1 Press the <CAL> key repeatedly until CELL appears on the
display.
2 Press the <RUN/ENTER> key.
3 Press the <CAL> repeatedly until the adjustable cell constant,
e.g. 0.278 cm
-1
appears.
4 Set the cell constant to be used with <V> <W> , e.g.
0.614 cm
-1
.