Manual
Operation handylab LF 11
24
ba75365e04 01/2011
Range
0.090 ... 0.110 cm
-1
5 To return to the measuring mode: Press the <M> key.
From now on, the cell constant 0.614 cm
-1
will be used.
1 Press the <CAL> key repeatedly until CELL appears on the
display.
2 Press the <RUN/ENTER> key.
3 Press the <CAL> key repeatedly until a cell constant in the
range 0.090 ... 0.110 cm
-1
appears on the display.
4 Set the cell constant to be used with <V> <W> , e.g.
0.105 cm
-1
.
5 To return to the measuring mode: Press the <M> key. Now this
setting is accepted for measuring.