4.4 Configuration
You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):
Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.
Reference
temperature
Reference temperature – 25 °C (TREF25)
– 20 °C (TREF20)
Display of the measured value as
conductivity or resistivity
S/cm or MΩ
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. t25 appears on the
display.
4 Using <V> <W>, switch between t25 and t20.
5 Confirm with <RUN/ENTER>. Cond appears on the display.