Manual

Operation handylab LF 11
32
ba75365e04 01/2011
4.5 Reset
Measurement
parameters
The following measured parameters (ϰ InI) are reset to the default condition:
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
Resetting the
measuring
parameters
Measuring mode
ϰ
Cell constant 0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
Temperature compensation nLF
Reference temperature
Temperature coefficient of the lin-
ear temperature compensation
2.000 %/K
TDS factor 1.00
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <CAL> key.
3 Using <V> <W>, switch between no and YES.
YES: Resetting the measuring parameters
no: Retaining settings.
4 Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the measuring
mode.