handylab LF 11 What to do if...
35
ba75365e04 01/2011
6 What to do if...
LoBat display
Instrument does
not react to key-
stroke
Error message OFL
Error message E3
Display to
Cause Remedy
Batteries almost empty Replace batteries
(see section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition
undefined or EMC load
unallowed
Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause Remedy
The measured value lies outside
the measuring range
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and
replace it if necessary
Unsuitable calibration
solution
Check calibration solutions
Cause Remedy
Time-out of the interface Check the instrument that is
connected