Manual
handylab multi 12 Operation
21
ba75371e03 11/2009
Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.
4.2.3 Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring in-
strument can measure the ORP voltage (U) of a solution.
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the <AR> key.
1 Perform the preparatory activities according to section 4.2.1.
2 Submerse the Redox electrode in the sample.
3 Press the <▲> <▼> keys until the U status display appears.
The ORP voltage (mV) of the sample appears on the display.
4 Wait for a stable measured value.