Probe evaluation After the calibration, the measuring instrument evaluates the current status
of the probe against the relative slope. The evaluation appears on the dis-
play. The relative slope has no effect on the measuring accuracy. Low values
indicate that the electrolyte will soon be depleted and the probe will have to
be regenerated.
Starting the
calibration
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet).
Display Relative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
E3
Eliminate the error according to
chapter 6 W
HAT TO DO IF...
S < 0.6 or S > 1.25
1 Perform the preparatory activities according to section 4.3.1.
2 Keep the air calibration vessel ready.
3 Put the D. O. probe into the air calibration vessel.
4 Press the <CAL> key repeatedly until the calibration mode ap-
pears. The sensor symbol displays the evaluation of the last
calibration (or no sensor symbol in the delivery state or after the
measurement parameter has been reset).