Manual

handylab multi 12 Operation
57
ba75371e03 11/2009
4.8 Configuration
You can adapt the handheld meter to your individual requirements. To do
this, the following parameters can be changed (the status on delivery is
marked in bold):
Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.
Baudrate
Baud rate 1200, 2400, 4800, 9600
Air pressure display Current value in mbar
(no input possible)
Calibration
intervals
pH 1 ... 7 ... 999 dInt 3
–O
2
1 ... 14 ... 999 dInt 4
ӕ: 1 ... 180 ... 999 dInt 5
Date/time Any
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
4 Select the required Baud rate with <> <>.
5 Confirm with <RUN/ENTER>. CAL dISP appears on the dis-
play.
Day.Month No.
Time
1
2
M
m
%
mg/l
cm
S/
cm
S
mbar
mV/pH
t
n
Store
Auto
LoBat
Tref25
nLF
/
K
%
RCL
Sal
°
AR
TP
C
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
OUpH
AutoCal DIN
AutoCal TEC
Ident
Year
0
0
8
4
Baud