Manual

Operation handylab multi 12
62
ba75371e03 11/2009
Resetting the
measuring
parameters
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <CAL> key. The setting to reset the pH measuring
parameters appears on the display.
3 Using <> <>, switch between no and YES.
YES: Resetting the pH measuring parameters
no: Retaining settings.
4 Confirm with <RUN/ENTER>.
The measuring instrument switches to the setting to reset the
O
2
measuring parameters.
5 Using <> <>, switch between no and YES.
YES: Resetting the O
2
measuring parameters.
no: Retaining settings.
6 Confirm with <RUN/ENTER>. The measuring instrument
switches to the setting to reset the Conductivity measuring pa-
rameters.