Manual

handylab multi 12 What to do if...
69
ba75371e03 11/2009
6.2 Oxi system messages
Error message 0FL
Error message E3
Error message E7
AR flashes
continuously
Cause Remedy
Display range exceeded
Oxygen sensor
Not connected Connect the probe
Cable broken Replace probe
Depleted Replace probe
Short-circuit between gold and
lead electrode
Clean probe and replace it if
necessary
Cause Remedy
Invalid calibration
Oxygen sensor
Electrolyte solution depleted Regenerate probe
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate probe
Worn out Replace probe
broken Replace probe
Cause Remedy
Membrane damaged
Membrane damaged Regenerate probe
Membrane head not screwed
on tight enough
Screw membrane head tight
Cause Remedy
No stable measured value
Membrane contaminated Clean membrane