Manual

handylab multi 12 What to do if...
71
ba75371e03 11/2009
6.4 General errors
Display LoBat
Instrument does
not react
to keystroke
Display to
Probe symbol
flashes
Message StoFull
Cause Remedy
Batteries almost empty Exchange the batteries (see
section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Hold the <CAL> key
depressed and switch the
instrument on
Cause Remedy
Time-out of the interface Check that the instrument is
connected
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
All memory locations are full Output data storage and clear
data storage