handylab Ox 12 Operation
25
ba75369e04 01/2011
4.3.1 Calibrating with the air calibration vessel
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet).
1 Connect the D. O. probe to the measuring instrument.
2 Keep an air calibration vessel ready for use.
3 Put the D. O. probe into the air calibration vessel.
4 Press the <CAL> key repeatedly until O
2
CAL appears.
5 Press the <RUN/ENTER> key. AutoRead is active, AR flashes.
6 As soon as a stable value is reached, the AR display indicator stops
flashing. This indicates that the calibration is complete; the sensor
symbol shows the relative slope that was determined and the sensor
evaluation (see page 24).