Instruction Manual

handylab Ox 12 What to do if...
49
ba75369e04 01/2011
6 What to do if...
LoBat display
Instrument does
not react to key-
stroke
Error message 0FL
Error message E3
Cause Remedy
Batteries almost empty Replace batteries
(see section 5.1 M
AINTENANCE)
Cause Remedy
Operating condition
undefined or EMC load
unallowed
Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause Remedy
Display range exceeded
D. O. probe:
Not connected Connect the probe
Cable broken Replace probe
Depleted Replace probe
Short-circuit between gold
and lead electrode
Clean probe and replace it if
necessary
Cause Remedy
Invalid calibration
D. O. probe:
Electrolyte solution depleted Regenerate probe
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate probe
Worn out Replace probe
broken Replace probe