What to do if... handylab Ox 12
50
ba75369e04 01/2011
Error message E7
AR flashes
continuously
Measured value
too low
Measured value
too high
Display
to
Probe symbol
flashes
Cause Remedy
Membrane damaged
Membrane damaged Regenerate probe
Membrane head not screwed
on tight enough
Screw membrane head tight
Cause Remedy
No stable measured value
Membrane contaminated Clean membrane
Cause Remedy
Insufficient
flow
Provide flow to the probe
Cause Remedy
High amount of dissolved
substances
Correct solubility function using
the salinity equivalent
Air bubbles bump on the
membrane with high velocity
Avoid direct flow to the
membrane
The carbon dioxide pressure
is too high (> 1 bar)
Measuring not possible
Cause Remedy
Time-out of the interface Check the instrument that is
connected
Cause Remedy
Calibration interval expired Recalibrate the measuring
system