Instruction Manual

Operation handylab pH 11
30
ba75366e04 01/2011
sponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the elec-
trode.
8 To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9 If necessary, enter the temperature of the second buffer solu-
tion with <> <>.
10 Immerse the pH electrode in the second buffer solution.
11 Press the <RUN/ENTER> key. The second pH value appears
on the display.
12 Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <> <>.
13 When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
14 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.