User guide
Settings pH 7310
44 ba75929e03 12/2011
7.3.2 Data storage
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see
section 8 DATA STORAGE, page 46).
7.4 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
7.4.1 Resetting the measurement settings
The following settings for pH measurements are reset to the default settings
with the Reset function:
System / Interface /
Output header
Output of a header for Output
format: CSV
System / Printer Settings for the integrated prin-
ter of the pH 7310 P.
For details, see section 10
PRINTER (ONLY PH 7310P),
page 54
System /Clock Date format
Date
Time
Time and date settings (see
section 4.4.6 EXAMPLE 2 ON
NAVIGATION: SETTING THE DATE
AND TIME, page 20)
System /Service
information
Hardware version and software
version of the meter are dis
-
played.
System /Reset - Resets the system settings to
the delivery condition (see
sec-
tion 7.4.2 RESETTING THE SYS-
TEM SETTINGS, page 45).
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting Default settings
Buffer AutoCalTEC
Cal. interval 7 d
Unit for slope mV/pH
Menu item Possible set-
ting
Description