pH 7310 What to do if...
ba75929e03 12/2011 59
12 What to do if...
12.1 pH/ORP
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
pH electrode:
Air bubble in front of the junction Remove air bubble
Air in the junction Extract air or moisten junction
Cable broken Replace electrode
Gel electrolyte dried out Replace electrode
Measured value outside the mea-
suring range of the meter
Use suitable electrode
Cause Remedy
pH electrode:
The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Recalibrate
Junction contaminated Clean junction
Electrode broken Replace electrode
Buffer solutions:
Wrong buffer solutions Change calibration procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions