User guide

What to do if... pH 7310
60 ba75929e03 12/2011
No stable measured
value
Obviously incorrect
measured values
Cause Remedy
pH electrode:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Temper if necessary
Electrode + test sample:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
pH electrode:
pH electrode unsuitable Use suitable electrode
Temperature difference between
buffer and test sample too great
Adjust temperature of buffer or
sample solutions
Measurement procedure not suit-
able
Follow special procedure