Owner manual

FDO
®
70x IQ (SW) What to do if...
6 - 1
ba76014e01 02/2012
6 What to do if...
Sensor does not appear
in the measured value
display and list of
sensors
Measured value
implausible
Measured value
fluctuating heavily
Cause Remedy
Sensor cap not mounted or
defective
Sensor cap (see section 5.3)
Replace a defective sensor
cap as necessary
Cause Remedy
Coating on sensor cap Clean the outside of the sen-
sor (see section 5.4.1)
Service life of the sensor cap
over
Check the sensor (see sec-
tion 4.2.2)
Replace the sensor cap as
necessary (see section 5.3)
Dirt inside the sensor cap and
sensor head.
Fixing ring not properly tight-
ened.
Sensor cap untight or defec-
tive.
Dismantle sensor cap
Clean the insides of the sen-
sor cap and sensor head (see
section 5.4.2)
Mount the sensor cap cor-
rectly and tighten the fixing
ring as far as it will go (see
section 5.3)
Replace a defective sensor
cap as necessary
Cause Remedy
Coating on sensor cap Clean the outside of the sen-
sor (see section 5.4.1)
Service life of the sensor cap
over
See point "Measured value
implausible"
Dirt inside the sensor cap and
sensor head.
Fixing ring not properly tight-
ened.
Sensor cap untight or defec-
tive.
See point "Measured value
implausible"