Owner manual
FDO
®
70x IQ (SW) Indexes
9 - 1
ba76014e01 02/2012
9 Indexes
9.1 Explanation of the messages
This chapter contains a list of all the message codes and related mes-
sage texts that can occur in the log book of the IQ S
ENSOR NET system
for the FDO
®
70x IQ (SW) sensor.
Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L
OG BOOK chapter of the IQ SENSOR NET sys-
tem operating manual.
Note
The last three digits of the message code identify the source of the
message:
334 = SC FDO 700 (sensor cap SC-FDO
®
700)
335 = SC FDO 701 (sensor cap SC-FDO
®
701)
531 = FDO700IQ (sensor basis / component class, adapter ADA)
9.1.1 Error messages
Message code Message text
EA133x Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA233x Sensor temperature too high!
* Check process and application
EA333x Sensor temperature too low!
* Check process and application
EAP33x Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap
EAP531 Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Replace sensor cap