User Manual

Lab 960 Operation
43
ba75557e05 11/2009
4.6.4 Erasing the data memory
You can erase all stored measurement datasets.
4.6.5 Downloading calibration data
You can download calibration data:
z to the display
via the memory menu
via the calibration menu
z to the interface (see page 45)
Download to display via
memory menu
1 Open the storage menu with <RCL>.
2 If necessary, scroll with <RCL> until Sto disp or Sto prt is
displayed.
3 Press <RCL_> to display the Clr All function.
4 Press <OK> to erase the entire contents of the memory.
or
Return to the measured value display with <MODE>.
The stored data is retained.
1 Open the storage menu with <RCL>.
2 If necessary, scroll with <RCL> until CAL diSP is displayed.
3 Press <OK> to display the calibration data.
The calibration date and calibrated cell constant are displayed
consecutively for approx. 10 seconds each.