User Manual
Operation Lab 960
56
ba75557e05 11/2009
4.9.2 Resetting meter settings
This function resets to the default condition meter settings. The
relevant values are given on the following pages:
The following settings are also reset to the default condition:
Resetting the meter
settings
System settings page 49
Measurement settings page 51
Setting Default settings
Measured variable
ӕ mS/cm or µS/cm
Adjusted cell constant 0,650 1/cm
Temperature compensation nLF
Temperature coefficient of the
linear temperature
compensation
2.000 %/K
TDS factor 1.00
1 Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2 During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3 Press <▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.