6.1 Error symptoms of D.O. measurement
Error symptom Cause Remedy
No D.O. display No connection between meter and
sensor
D.O. module not mounted
properly
Cable defective
Check connection between meter
and sensor
Check the plug-in connection,
clean the contacts as necessary
Return the sensor
The sensor is in the air
and the display shows
0.0 mg/l or 0 % O
2
No connection between meter and
sensor
D.O. module not mounted
properly
No electrolyte in the membrane
cap
Cable defective
Check connection between meter
and sensor
Check the plug-in connection,
clean the contacts as necessary
Replace and refill the membrane
cap (see section 5.3)
Return the sensor
The sensor cannot be
calibrated
Contaminated membrane cap
Electrolyte depleted
Membrane damaged
Replace and refill the membrane
cap (see section 5.3)
Subsequently, wait for 30 to
50 min and recalibrate.
The sensor still cannot
be calibrated after
changing the electrolyte
and membrane cap
Contaminated electrodes or
sensor toxification
Membrane cap not screwed on
tight enough
Clean the electrodes (see
section 5.4)
Screw membrane cap tighter
Incorrect temperature
display
The temperature sensor of the
conductivity measuring cell is not
immersed in the measuring
solution
Temperature sensor defective
Observe the minimum immersion
depth
Return the sensor
Measured values implau-
sible
Incident flow insufficient Move the sensor in the test
sample (stirr) or use the battery-
powered stirrer, BR MPP 350
Mechanical damage to
the sensor
Return the sensor