User Manual

MPP 350 What to do if...
23
ba75434e02 07/2005
6.2 Error symptoms of conductivity measurement
6.3 Error symptoms of pH measurement
Error symptom Cause Remedy
No temperature or con-
ductivity display
No connection between meter and
sensor
Cable defective
Check connection between meter
and sensor
Return the sensor
Measurement delivers
implausible conductivity
values
Incorrect cell constant adjusted at
the meter
Measuring range exceeded
Contamination in the area of the
electrodes
Electrodes damaged
Check / correct the cell constant
Observe the application range
Clean the sensor (see
section 5.2).
Return the sensor
Incorrect temperature
display
The temperature sensor was not
immersed in the measuring
solution
Temperature sensor defective
Observe the minimum immersion
depth
Return the sensor
Error symptom Cause Remedy
No pH display No connection between meter and
sensor
Cable defective
pH electrode defective
Check connection between meter
and sensor
Return the sensor
Replace pH electrode