Commissioning MPP 350
8
ba75434e02 07/2005
Note
The membrane cap that is mounted on the D.O. sensor for delivery serves
mainly as a transport protection. Depending on the duration of the transport
and storage period, it may have a shortened operational lifetime. If the mea-
suring system cannot be calibrated (error message on the instrument),
please proceed according to section 5.3 D.O.
MODULE: CHANGING THE ELEC-
TROLYTE SOLUTION AND MEMBRANE CAP.
3.1.2 Set equipment
The MPP 350 is also available as a set in the MPP-FC field case. You can
also buy the carrying case as an accessory and fit it with components your-
self. Electrodes with a cable length of up to 25 m fit in the carrying case. Or-
dering information, see chapter 8 W
EAR PARTS AND ACCESSORIES.
Equipment of the
field case
MPP-FC
(example)
1 Electrode or battery-powered
stirrer
8 Reagents (buffers etc...)
2 Manual, CD-ROM 9 Plug
3 Meter 10 Protective hood with sinker
4 Plug-in power supply unit 11 Exchange membrane caps
1
2
4
5
67
8
9
10
13
11, 12
8
8
3