6.2 Oxi system messages
Error message 0FL
0FL0FL
0FL
Error message E3
E3E3
E3
Error message E7
E7E7
E7
AR flashes
continuously
Cause Remedy
Display range exceeded
Oxygen sensor
Not connected Connect the sensor
Cable broken Replace sensor
Depleted Replace sensor
Short-circuit between gold and
lead electrode
Clean sensor and replace it if
necessary
Cause Remedy
Invalid calibration
Oxygen sensor
Electrolyte solution depleted Regenerate sensor
Membrane contaminated Clean membrane
Electrode system poisoned Regenerate sensor
Obsolete Replace sensor
Broken Replace sensor
Cause Remedy
Membrane damaged
Membrane damaged Regenerate sensor
Membrane head screwed on
not tight enough
Screw membrane head tight
Cause Remedy
No stable measured value
Membrane contaminated Clean membrane