Manual
Multi 1970i What to do if...
69
6.2 Oxi system messages
Error message 0FL
0FL0FL
0FL
Error message E3
E3E3
E3
Error message E7
E7E7
E7
AR flashes
continuously
Cause Remedy
Display range exceeded
Oxygen sensor
– Not connected – Connect the sensor
– Cable broken – Replace sensor
– Depleted – Replace sensor
– Short-circuit between gold and
lead electrode
– Clean sensor and replace it if
necessary
Cause Remedy
Invalid calibration
Oxygen sensor
– Electrolyte solution depleted – Regenerate sensor
– Membrane contaminated – Clean membrane
– Electrode system poisoned – Regenerate sensor
– Obsolete – Replace sensor
– Broken – Replace sensor
Cause Remedy
Membrane damaged
– Membrane damaged – Regenerate sensor
– Membrane head screwed on
not tight enough
– Screw membrane head tight
Cause Remedy
No stable measured value
– Membrane contaminated – Clean membrane