User Manual

Operation Multi 3410
42
ba75863e11 02/2011
Calibration /Unit for slope mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst
slope).
QSC /First calibration - Starts the initial calibration with QSC buffers.
This menu item is only available as long as
no initial calibration was carried out with the
connected IDS sensor.
QSC /Record of first calibration - Displays the calibration record of the QSC ini-
tial calibration
QSC /Control calibration - Starts the control calibration with QSC buf-
fers.
This menu item is only available if an initial
calibration was carried out with the con-
nected IDS sensor.
Man. temperature -25 ... +25 ...
+130 °C
Entry of the manually determined tempera-
ture
This menu item is only available if an IDS
adapter and is connected.
pH resolution 0.001
0.01
0.1
Resolution of the pH display
mV resolution 0.1
1
Resolution of the mV display
Reset - Resets all sensor settings to the delivery con-
dition (see section 4.12.1)
Menu item Possible setting Description