6.2 pH
Error message,
OFL, UFL
Error message,
Error
Cause Remedy
IDS pH sensor:
Measured value outside the
measuring range
Use suitable IDS pH sensor
Air bubble in front of the junction Remove air bubble
Air in the junction Extract air or moisten junc-
tion
Cable broken Exchange IDS pH sensor
Gel electrolyte dried out Exchange IDS pH sensor
Cause Remedy
IDS pH sensor:
The values determined for zero
point and slope of the IDS pH
sensor are outside the allowed
limits.
Recalibrate
Junction contaminated Clean junction
IDS pH sensor broken Exchange IDS pH sensor
Buffer solutions:
The used buffer solutions are not
suitable for the selected buffer
set
Select a different buffer set
or
Use different buffer solu-
tions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions