User Manual
Operation Multi 3500i
32
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see
section 4.8.5). Subsequently, you can transmit the displayed calibra-
tion data to the interface, e. g. to a printer or PC, with the <PRT> key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record:
03.11.03 07:14
CALIBRATION pH
03.11.03 07:10:45
Multi 3500i ser. no. 12345678
Cal. interval 7 d
AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 0.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope -60.2 mV/pH
Asymmetry 3.0 mV
Sensor +++