Multi 3500i Operation
45
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record:
Calibration evaluation After calibrating, the measuring instrument automatically evaluates the
calibration.
03.11.03 07:14
CALIBRATION ISE
03.11.03 07:12:01
Multi 3500i ser. no. 12345678
Standard 1 0.010 mg/l
Standard 2 0.020 mg/l
Voltage 1 0.0 mV 24.0 °C
Voltage 2 9.0 mV 24.0 °C
Slope 29.9 mV
Sensor +++
Display Calibration
record
Magnitude of the slope [mV]
+++ 50.0 ... 70.0 or 25.0 ... 35.0
Error Error
< 50 or > 70
or
< 25 or > 35
Perform error elimination ac-
cording to chapter 6 W
HAT TO
DO IF...